4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 86
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.2.12
Test 1.12 Concentrate Pumps and Conductivity
1.
Activate the Concentrate Pumps and
Conductivity menu with the Concentrate
Pumps and Conductivity key in the LL Manual
Test menu. The following menu is displayed.
2.
Check the functions with the appropriate keys.
You can skip directly to the calibration of the conductivity sensors
BICLF/ENDLF
by pressing the
key.
You can set the pumps
BICP/KP
, flow pump
FPE
and the heater power
H %
.
All balance chamber valves are open and the flow pump
FPA
works with the
same speed as
FPE
.
The machine is in main flow. The machine can suck-in concentrate with the
concentrate pumps
BICP/KP
for the preparation of the conductivity to test the
conductivity sensors
BICLF/ENDLF/ENDLF-S
.
You can connect a measurement instrument at the dialyser couplings or
between
FPE
and the balance chamber.
The supervisor displays the head speed
BICP-S/KP-S
of the concentrate pumps.
Increase the speed [rpm] until the flow pump
FPE
has built up a flow of 750
ml/min. Thereby the pulsating flow of the pumps have no influence on the
conductivity measurement.