4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 98
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.2.20
Test 1.20 Water Part (Overview)
1.
Activate the Water Part (Overview) menu with
the Water Part (Overview) key in the LL
Manual Test menu. The following menu is
displayed.
2.
Check the functions with the appropriate keys.
You can set the speed for the flow pump
FPE/FPA
. The heater power
H %
can
be set in %. The speed of the concentrate pumps
BICP/KP
can be set manually.
After starting this menu the balance chamber is initialised. Then the flow pump
FPE
is set to the same speed as
FPA
. The balance chamber valves are switched
automatically.
Main Flow
The machine can be switched between main flow and bypass for servicing.
Main Flow
Valves
VDA/VDE
are open (valve symbols are black).
Valve
VBP
is closed (valve symbol is white).
Bypass
Valves
VDA/VDE
are closed (valve symbols are white).
Valve
VBP
is open (valve symbol is black).