3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 55
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.5.8
Self Test
1.
Turn service switch
S1
to position 0 - Therapy.
2.
Start machine (fully equipped) in dialysis mode.
3.
Insert line system – see instructions for use.
4.
The machine must run without errors up to the message Connect Patient.
3.5.9
Test Run
5.
Assemble the blood lines, perform preparation (insert heparin syringe and
select 2 ml/h and use BIC cartridge if present).
6.
Perform self test – according to instructions for use.
7.
Press the
icon to switch to therapy mode (connect patient). This
icon is enabled after all self tests were performed successfully.
8.
Insert the arterial and venous lines in the graduated cylinder.
9.
Set the blood pump so that no blood side alarms are activated.
10.
Start dialysis.
11.
Check alarm function of SAD.
(The ABPM function (if present) can be checked without a test run of the
machine in preparation mode, without the concentrate pump.)
12.
Terminate dialysis.
13.
Perform disinfection/decalcification.
14.
Remove blood lines, filter and heparin syringe.
15.
Switch off .