3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 69
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.6.7.3 Write System Configuration
The utility writes the System Configuration from the USB stick to the hard disk
drive (CFC) of the machine. The system configuration must be read to
the USB stick with the 10 Read System Configuration utility, because by default
the directory is empty.
System Configuration
The system configuration saved on the FSU USB stick must be downloaded
to an other machine only if:
•
the hardware matches and
•
the identical TLC software version number is present.
The last write procedure can be undone with the
07 Undo Last Written
utility.
USB Stick with Treatment Support Settings
Old treatment support settings must not be downloaded via a USB stick after a
software update. If necessary download the new treatment support setting
after an update.
Activation of the Treatment Support Settings
The settings are only activated after the is switched off and on again.
Check the settings after the was switched on again.
BlueCat
Removable Disk (X: - FSU USB Stick)
DISKTYPE
USER
05_Write_Calibration_Data
DEFAULT
04_Write_Languages
06_Write_System_Configuration
If system configuration is not on the FSU USB stick
The system configuration must be read to the USB stick with the
10 Read System Configuration utility, because by default the
directory is empty.
Write System Configuration
USB:\USER\06_Write_System_Configuration\DEFAULT\config.xxu
Note:
The System Configuration file can be copied back to the
machine only, if the TLC version number, saved in
the System Configuration file, is equal with the TLC
version number reported by the machine.
F i e l d S e r v i c e U t i l i t i e s V 2 . 0 0 - S e t M S T
CRC = XXXX - OK
01. Set System Version Number
02. Set Working Time Counter (WTC)
03. Kill Masterboot Record
( Hard Disk)
04. Write Languages
(USB Stick
→
Hard Disk)
05. Write Calibration Data
(USB Stick
→
Hard Disk)
06. Write System Configuration
(USB Stick
→
Hard Disk)
07. Undo Last Written
(USB Stick
→
Hard Disk)
08. USB Stick Self Test
09. Select Menu of Card Reader Commands
10. Read System Configuration
( Hard Disk
→
USB Stick)
11. Read Calibration Data
( Hard Disk
→
USB Stick)
12. Read All Trends
( Hard Disk
→
USB Stick)
13. Read Screenshots
( Hard Disk
→
USB Stick)
14. Get Error (ErrorDisk)
( Hard Disk
→
USB Stick)
BP-
: move cursor up
BP+
: move cursor down
AQ
: exit – see
Important Note
EQ
: execute
Important Note:
To prevent the corruption of USB stick, before removing it,
please press
AQ
then follow the instructions on the screen!
1.
Press the
BP
hardware key and move the cursor
down to select the utility.
2.
Press the
EQ
hardware key to execute the
utility.