4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 161
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
•
Manometer 0 to 4 bar (art. no. 7701357 or equivalent)
•
Flow meter (art. no. 770085A or equivalent)
•
Allen key 3 mm
•
Size 7 spanner
Please ensure that the manometer and connection tubing are at the same
level as the nonreturn valve during calibration.
1.
Connect manometer and flow meter between flow pump
FPE
and
FBK1
.
2.
Select Test 1.20 Water Part (Overview) in TSM.
3.
Increase slowly the speed of
FPA
, until a flow of 500 ml/min is reached.
4.
Set temperature
TSD
to 37
o
C and wait until temperature is reached.
5.
Clamp tubing between manometer and
FBK1
.
6.
Loosen lock nut on
RVFPE
adjustment screw.
7.
Set a pressure of 975 mmHg (1300 mbar) with the adjustment screw of
RVFPE
.
•
975 mmHg
±
37 mmHg
•
1300 mbar
±
50 mbar
8.
Tighten lock nut again.
9.
Remove manometer and flow meter.