4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 169
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.25
Gap Arterial Tubing Clamp SAKA/Venous
Tubing Clamp Currentless Closed SAKV-SG
You can calibrate the gap for the arterial tubing clamp
SAKA
and
the venous tubing clamp currentless closed
SAKV-SG.
•
Calibration template 1.4/1.5 (art. no. 7702493)
1.
Remove cap
C
(e.g. use a sharp knife to lift out).
2.
Close the respective tubing clamp in Manual Test and
Calibration, LL Manual Test, Test 1.5 Tubing Clamps.
Fig.: Tubing Clamp SAKV-SG
Fig.: Tubing Clamp SAKA
Set gap
G
to 1.4 mm with 1.4 mm template.
3.
Loosen screw
F
.
4.
Push 1.4 mm template between clamp
D
and eccentric cam
E
.
5.
Rotate eccentric cam
E
until 1.4 mm is set.
6.
Tighten again screw
F
Note
The 1.4 mm template fits through the gap.
7.
Check wit the 1.5 mm template.
Note
The 1.5 mm template must not fit through the gap.
8.
Reassemble cap
C
.