4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 128
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.7
Calibration Degassing Temperature Sensor TSE/
Heater Inlet Temperature Sensor TSHE
Prior to Calibration
•
Position the heater inlet temperature sensor
TSHE
next to
TSE
(see figure
redirect heater inlet temperature sensor
TSHE
, next page)
•
Use calibrated measurement instrument
•
Wait until Dialog has reached working temperature
The degassing temperature sensor
TSE
and the heater inlet temperature sensor
TSHE
are calibrated together. The following calibration points are used:
TSE
/
TSHE
:
30 °C, 40 °C, 85 °C
EK
DMV
VVBE
TSE
PE
TSHE
VVB*
VBKO*
DBK*
T
A
D
T
RVBO*
VBKS
TSBIC
VBICP*
SKBS
RVDA
VLA
H
NS
V
B
VB WT
FEP
WAB
EP
VEB
TSH
T
8
. ..
EK
DMV
VVBE
PE
TSHE
TSHE
VVB*
VBKO*
DBK*
T
A
D
T
T
RVBO*
VBKS
H
NSV
B
VB WT
FEP
WB
EP
TSH
T
8
Measurement
Instrument
SKBS
C
mS/cm
o
ml/min
VEB
TSE
Fig.: Temperature Circulation
Fig.: Loop-In Point and TSHE in Line with TSE for Calibration
EP
RVDA
WAB
H
WT
VEB
KP/BICP
FEP
TSE
NSVB
TSHE
DBK
EP
RVDA
H
WT
VEB
KP/BICP
FEP
TSE
NSVB
C
mS/cm
o
ml/min
Measurement
Sensor
Keep tubings as short
as possible
Measurement
Instrument
WAB
DBK
TSHE
Fig.: Original Tubing Water Sub-Rack for TSE/TSHE
Fig.: Water Sub-Rack with TSHE in Line with TSE for Calibration