4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 82
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.2.8
Test 1.8 Coupling Detectors
1.
Activate the Coupling Detectors menu with
the Coupling key in the LL Manual Test menu.
The following menu is displayed.
2.
Check the functions of the reed contacts with
the appropriate couplings.
You can check the actual status of the reed sensors.
Detectors
BICSS
Bicarbonate rinsing connection sensor.
KSS
Concentrate rinsing connector sensor.
SBS1
Rinsing bridge connector sensor (dialyser coupling).
SBS2
Rinsing bridge connector sensor (dialyser coupling).