4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 114
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.2
Select Low Level Manual Calibration
You can calibrate all sensors in the LL Manual
Calibration menu. The number of calibration
points is preset.
Some sensors have the possibility to run a
combined or a single calibration.
The TSM Main Menu is displayed after the
machine is switched on (set service switch S1 to
position 2, on digital board).
1.
Activate the Manual Test & Calibration menu
with the Manual Test & Calibration key in the
TSM Main Menu. The following menu is
displayed.
2.
Activate the LL Manual Calibration menu with
the LL Manual Calibration key in the Manual
Test and Calibration menu. The following menu
is displayed.
The following low level calibrations can be
selected:
•
2.1 Pressure Sensors (Blood Side)
•
2.2 Heparin Pump (not applicable)
•
2.3 Pressure Sensors (Water Side)
•
2.4 Degassing Temperature
•
2.5 Dialysis Temperature
•
2.6 Conductivity
•
2.7 Blood Leak Detector
•
2.8 Flow of BICP, KP and UF
•
2.9 Safety Air Detector SAD
•
2.10 BICLF Ratio and ENDLF Ratio
•
2.11 Load Cell (not applicable)
•
2.12 PV Alarm Window, PA Low Limit
•
2.13 Flow of BPA and BPV Pump
•
2.14 Substitution Pump (not applicable)
•
2.15 HDF Online Substitution Pump OSP
•
2.16 Level Regulation