4.2
Quit TSM Service Program
TSM Service Program
Only activate the TSM service program for
service activities. It is prohibited to connect a
patient to the and to run a therapy if
the TSM service program is activated in the
. If the TSM service program is
activated the complete alarm system is
disabled.
Therapy Mode
After completion of all procedures switch back
to the therapy mode: digital board, service
switch S1, position 0.
The following steps must be taken to quit the
service program and to re-enter the therapy
mode:
1.
Switch off machine.
2.
On digital board DB: turn service switch S1
from position 2 to
position 0
therapy mode.
Service Switch S1:
Position 0
= Therapy
Position 2
= TSM Service Program
Position 3
= Software Installation/Update
3.
Close rear door.
4.
Switch on machine.
The machine starts in therapy mode.