Test 1.13 Balance Chamber Valves
1.
Activate the Balance Chamber Valves menu
with the Balance Chamber Valves key in the LL
Manual Test menu. The following menu is
displayed.
2.
Check the functions with the appropriate keys.
You can activate a single valve or combined setting via the controller. The
actual condition is displayed by the supervisor. The following valves can be
activated:
Valves
VDEBK1/2
Dialyser inlet valve 1/2 balance chamber
VEBK1/2
Inlet valve 1/2 balance chamber
VDABK1/2
Dialyser outlet valve 1/2 balance chamber
VABK1/2
Outlet valve 1/2 balance chamber