3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 104
BA-TE-DE08C M.KAY
Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
SW9.xx
3.8.2
Tests and Measures to be Executed
It is recommended to use the repair matrix, the tests and measures and the
execution protocol after servicing (e.g. repair, SW installation etc.).
All components are numbered alphabetically (from 1 to n) in the column Points
for Exchanged/Repaired Assembly Groups. The respective calibrations, tests and
measures in the matrix are assigned to each component (HW/SW) with a
◆
character. The executed measure can be documented in the execution protocol.
Risk of a patient infection due to a contamination of the manometer
protection filter of the blood line system!
If the manometer protection filter of the blood line system was
contaminated with blood and blood has entered the machine: replace the
internal manometer protection filter, the tubing (between the filter and
the manometer connector) and the manometer connector.
Tubing must be replaced only by the same tubing type/length and identical
installation manner.
Make sure that the tubings in the machine are not kinked or twisted after
servicing (e.g. if sub-racks are pulled out and inserted again). The tubing
must not touch moving/rotating components (e.g. motors of gear pumps).
3.8.2.1
Install Software
1.
The software must be installed according to the description in the service
manual.
3.8.2.2
Check System Configuration, Production
Report and Set if Necessary
The FSU USB stick with the system configuration can only be used if the
hardware and software of the machine match.
1A
System Configuration
•
System Configuration Present
Load the system configuration from the FSU USB stick and check all
settings in the TSM service program.
•
Customer Specific System Configuration not Present)
Check system configuration and set system configuration according to
user default.
1B
Production Report
•
Select and save/activate present options.
1C
Calibration Data
•
Replacement of CFC
Save calibration data.
•
Replacement of Sensors/Detectors
Calibration, save calibration data.
•
Replacement of Digital Board
Download calibration data.
•
Replacement of Analog Board
Calibration, save calibration data.
1D
LLC Calibrations
Check and set ratio, tubing constants BPA/BPV, OSP tubing type,
check/set PA/PV limit values.
1E
Working Time Counter
•
Set working time counter.
1F
Motherboard
Check and set BIOS.