4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 156
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.16
Calibration Flow Rate of BPA and BPV/OSP
The tube constants can be directly entered or calculated via a comparison measurement.
BPA
and
BPV
/
OSP
(option HDF online substitution pump) can be activated with the
START Calibration key. The delivered volume is measured in ml and entered as reference
value. The pressure conditions for the tube system during measurement must be
identical with the pressure conditions in therapy.
1.
Activate the LL Manual Calibration menu with
the LL Manual Calibration key in the Manual
Test and Calibration menu.
2.
Activate the Flow of BPA and BPV/OSP menu
by pressing on the line Calibration 2.13 Flow of
BPA and BPV/OSP. The following menu is
displayed.
Tube Constant
1.
Press the Tube Constant key to activate the
menu Calibration 2.13.1 Tube Constants of BPA
and BPV. The following menu is displayed.
The following tube constants must be entered for
the applied tubing system.
8 x 12 Tubing System
BPA
:
12306 µl/head rotation
BPV
:
12306 µl/head rotation
7 x 10 Tubing System
BPA
:
9293 µl/head rotation
BPV
:
9296 µl/head rotation
Pump Head
1.
Press the Pump Head key to activate the menu
Pump Head Selection. The following menu is
displayed.
You can select the pump head roller type used in
the blood pumps.
•
Pump head 8 x 12
•
Pump head 7 x 10 (option)