3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 61
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.6.5
Set Working Time Counter (WTC)
The Working Time Counter for the can be set with this utility. The
utility is typically used after replacing a hard disk drive (CFC). There is no file
exchange between the USB stick and hard disk drive (CFC). In case of
acceptance the new value is written to the hard disk drive (CFC) (TLC
WTC files).
F i e l d S e r v i c e U t i l i t i e s V 2 . 0 0 - S e t M S T
CRC = XXXX - OK
01. Set System Version Number
02. Set Working Time Counter (WTC)
03. Kill Masterboot Record
( Hard Disk)
04. Write Languages
(USB Stick
→
Hard Disk)
05. Write Calibration Data
(USB Stick
→
Hard Disk)
06. Write System Configuration
(USB Stick
→
Hard Disk)
07. Undo Last Written
(USB Stick
→
Hard Disk)
08. USB Stick Self Test
09. Select Menu of Card Reader Commands
10. Read System Configuration
( Hard Disk
→
USB Stick)
11. Read Calibration Data
( Hard Disk
→
USB Stick)
12. Read All Trends
( Hard Disk
→
USB Stick)
13. Read Screenshots
( Hard Disk
→
USB Stick)
14. Get Error (ErrorDisk)
( Hard Disk
→
USB Stick)
BP-
: move cursor up
BP+
: move cursor down
AQ
: exit – see
Important Note
EQ
: execute
Important Note:
To prevent the corruption of USB stick, before removing it,
please press
AQ
then follow the instructions on the screen!
1.
Press
BP
and move the cursor down to select
the utility.
2.
Press
EQ
to execute the utility.
D i a l o g + W o r k i n g T i m e S e t t i n g
- - - - - - - - - - - - - - - - - - - - - - - - - - - -
If you want to set the WTC create the new value with the HW-keys.
I f n o t , e x i t w i t h t h e A Q k e y .
N e w v a l u e f o r W o r k i n g T i m e C o u n t e r : 0 0 0 0 0
BP-
: decrement digit
BPstart
: next digit
BP+
: increment digit
AQ
: cancel
EQ
: accept
An input screen opens to set the new Working Time
Counter (WTC) value with the following hardware keys:
BP
:
decrements the current digit
BP
:
increments the current digit
BP
stop
stop
:
selects the next digit (from right to left)
AQ
: cancels the operation, WTC is not changed
EQ
: sets the new WTC value
3.
Set the new working time counter with the
BP
,
BP
and
BP
stop
stop
hardware keys.
4.
Press
EQ
to accept the entered value.