4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 168
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
•
Manometer 0 to 2.5 bar (art. no. 34511571 or equivalent)
The calibration of
DMV
is performed in the dynamic operating mode (
FPE
running, level in the upline tank is controlled).
1.
Disconnect water inlet hose from wall connection (close water tap if
necessary).
2.
Start briefly the unit to achieve a pressure free condition in the water
inlet part to avoid water spraying.
3.
Remove locking screw from
DMV
and screw in the manometer (if
necessary use teflon tape to prevent leakages).
4.
Connect water tubing and open the water tap.
5.
Open the following menus in the TSM service program: Manual Test and
Calibration, LL Manual Test and 1.9 Water Inlet, Upline Tank and Flow
Pump FPE.
6.
Set
FPE
to approx. 1400 rpm (approx. 1200 ml/min) and open
VVBE
(the
level in the upline tank must be controlled between medium and high
level).
7.
Pull up the adjustment knob to unlock pressure reducer valve
DMV
and
set pressure to approx. 0.9 ±0.1 bar with the adjustment knob.
8.
Close water tap.
9.
Press down the adjustment knob to lock pressure reducer valve
DMV
.
10.
Switch off Dialog.
11.
Disassemble manometer.
12.
Reassemble locking screw.