4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 88
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.2.14
Test 1.14 Bypass and Disinfection Valves
1.
Activate the Bypass and Disinfection Valves
menu with the Bypass and Disinfection Valves
key in the LL Manual Test menu. The following
menu is displayed.
2.
Check the functions with the appropriate keys.
You can activate a single valve or combined setting via the controller. The
actual condition is displayed by the supervisor. The following valves can be
activated:
Valves
VBP
Bypass valve
VDE
Dialyser inlet valve
VDA
Dialyser outlet valve
VD
Disinfection
valve
VZ
Circulation
valve
VLA
Air separator valve