4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 119
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.5
Calibration Degassing Pressure Sensor PE
Prior to Calibration
•
Use calibrated manometer (or pressure measurement instrument)
•
Position the manometer at the same level (see
A
in bottom figure) as the
degassing pressure sensor
PE
•
Keep the extra tubings as short as possible for calibration
•
Wait until Dialog has reached working temperature
You can calibrate the degassing pressure sensor
PE
. Build up the negative
pressure in the degassing circuit with a syringe.
EK
DMV
VVBE
TSE
VVB*
VBKO*
DBK*
T
RVBO*
VBKS
TSBIC
VBICP*
SKBS
RVDA
VLA
H
NS
V
B
VB WT
WAB
TSH
T
PE
TSHE
A
D
T
FEP
EP
VEB
8
A
Fig.: Degassing Circuit and PE