3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 67
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
F i e l d S e r v i c e U t i l i t i e s V 2 . 0 0 - S e t M S T
CRC = XXXX - OK
W r i t e C a l i b r a t i o n D a t a ( U S B S t i c k - > D i a l o g + H a r d D i s k )
N o s o u r c e i s f o u n d . . .
AQ
: back to previous menu
The utility searches for sources. An input screen can
open with the following information (see left figure),
i.e. no sources were found on the USB stick because the
calibration data subdirectories are empty. Calibration
data must be read to the USB stick before you can
write calibration data to the hard disk drive
(CFC).
3.
Press
AQ
to go back to the previous menu.
4.
Use the Read Calibration Data utility to read the
calibration data from the hard disk drive
(CFC) to the USB stick.
F i e l d S e r v i c e U t i l i t i e s V 2 . 0 0 - S e t M S T
CRC = XXXX - OK
W r i t e C a l i b r a t i o n D a t a ( U S B S t i c k - > D i a l o g + H a r d D i s k )
Writing Calibration Data to from: Ser-No./SYSver=90832/VX_XX
A r e y o u s u r e ?
AQ
: no
EQ
: yes
An input screen opens to Write Calibration Data to the
hard disk drive (CFC) with the following hardware keys:
AQ
: No, calibration data are not written to
hard disk drive (CFC), exit to main
menu
EQ
:
Yes, calibration data are written to
hard disk drive (CFC)
Ser-No./SYSver=90832/VX_XX
The utility compares the serial number (Ser-No.) and the system version number
(SYSver) stored in the calibration data file with the serial number and the
system version number from the . The calibration data file can be
written only if they match.
Example
Ser-No.= 90832
SYSver = V8_20