4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 102
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.2.24
Test 1.24 HDF Online
Exchange Intervals:
The HFB filter must be exchanged on an annual basis.
1.
Activate the HDF Online menu with the HDF
Online key in the LL Manual Test menu. The
following menu is displayed.
2.
Check the functions with the appropriate keys.
You can check the function of the HDF online substitution pump OSP (BPV) and
the additional valves.
Online Substitution Pump
OSP (BPV)
The function of the HDF online substitution pump OSP (BPV) can be checked.
Valves
VBE, VDFF, VSB, VSAE, VSAA
The flow path through both filters (
DF
and
HDF
) can be set via the valves.
Before operation in main flow the DF filter inlet must be vented via
VDFF
,
because the membranes of the
DF
filter and
HDF
filter are impermeable to air.
Valve
VBE
can only be opened if
PDA
< -100 mmHg to prevent fluid
penetration.