3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 7
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.2.1
Set-Up Low Level Controller
LLC Test Codes
Format of LLC Test Codes:
T T
S S
x:
T T
S S
x
LLC test code to LLS
Performed test step
Number of performed repeats
0
≤
x < repeats until error message
Examples
11102
Third blood leak detector test: send signal to blood leak detector task
33701
Second BS pressure test terminated, because during autopriming, e.g. an LF
error occurred.
37991
VD test passed after the second time.
LLC Test Codes Before the
Individual Tests
T T
00
x
Start test.
T T
01
x
Error: timeout (10 s) at "Before Test Start Wait for Test Code 0 from LLS".
T T
02
x
Test check prerequisites (if present, see respective tests).
T T
03
x
Error: Check timeout test prerequisites (if present, timeout time see respective
tests).
T T
04
x
Wait for LLS test code from LLS.
T T
05
x
Error: timeout (10 s) at "Before Test Start Wait for LLS Test Code from LLS".
Test Codes of the Individual Tests
Range 10x .... 99x
Not all test codes are displayed sequentially, due to the implementation.
A higher test code always includes all the finalised subcodes with lower
priority.
In case of an error: The subcode always includes the subphase where the error
occurred.
3.2.2
Set-Up Low Level Supervisor LLS
Test Codes
The test code is displayed as a numerical value
≥
10000. If the value is < 9000,
this is the required rinsing volume for the HDFO filter rinsing in ml. If the value
is > 9000 and < 10000, the last three digits display the required HDFO filter
rinsing volume in ml through the valves VSAA and VSAE (VSB is closed).
Format of LLS Test Codes:
T T
S S S
:
(Value
≥
10000)
T T
S S S
LLS test code to LLC
Performed test step
Example:
11104
Blood leak detector test: measure level for red and green