4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 141
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
FVD
blue
red
Fig.: Red and Blue Couplings Connected to Rinsing Bridge
FVD
blue
red
Fig.: Rinsing Bridge with Red Coupling Immerged in Reference Solution for BL Calibration
1.
You can activate the LL Manual Calibration
menu with the LL Manual Calibration key in
the Manual Test and Calibration menu.
2.
Activate the Blood Leak menu by pressing on
the line Calibration 2.7 Blood Leak. The
following menu is displayed.
Close front and rear door for the function test and calibration to prevent light
irradiation on the sensor.
Leakages
Air and micro-bubbles can lead to incorrect measurement values in the
measurement line and can cause fluctuation of the current value.
If air is visible check the complete system for leaks in menu 1.20 Test Water
Part. After the initialisation phase rinse the blood leak detector approx. 1 min
and then perform a leakage test of the system in menu 1.19 Leak Test.
Chemical Disinfection with Citric Acid 50%
A decalcification must be performed before initial calibration.