4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 44
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Central Automatic Chemical
Inlet Volume
The inlet volume can be selected.
Switch-off for retention without automatic
switch-on
NO – (i.e.: the automatic switch-on of the
machine is performed)
The retention time of the disinfectant in the
machine is according to the preset retention time.
YES – (i.e.: the automatic switch-on of the
machine is not performed)
The retention time of the disinfectant in the
machine depends on the switch-off time and from
the manual switch-on of the machine by the user.
Retention Time
If NO was selected
The retention time is enabled and can be selected.
If YES was selected
The retention time is disabled and can not be
selected. The retention time must be
≤
10 hours
for manual retention.
Rinsing Flow
The rinsing flow can be selected.
Rinsing Time
The rinsing time can be selected.
Central Chemical
A central chemical mode can be selected for the
automatic switch-on in the weekly schedule.