4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 9
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.4
Service Overview in Therapy
The service overview can be selected in therapy with the Service Key icon. The
second and third page of the service overview can be selected with the service
key icon.
Click the Door icon to jump back to the initial menu.
W a t e r T a n k
VVBE: OPENED/CLOSED
NSVB: HIGH/LOW
D e g a s s i n g
EP: XXXX
rpm
VEB: OPEN/CLOSED
PE: XXX
mmHg
TSE: XX.X
o
C
TSHE: XX.X
o
C
H e a t e r
HEATER: XX.X
%
A i r S e p a r a t o r
LAFSO: NO
AIR/AIR
LAFSU: NO
AIR/AIR
VLA: OPEN/CLOSED
T e m p e r a t u r e
TSE: XX.X
o
C
TSBIC: XX.X
o
C
TSD: XX.X
o
C
TSD_S: XX.X
o
C
TSDE: XX.X
o
C
C o n d u c t i v i t y
BICLF: X.X
mS/cm
ENDLF: XX.X
mS/cm
ENDLF_S: XX.X
mS/cm
BICP: XX
rpm
BICP_S: XX
rpm
BIC-Ratio: XX.X
BIC-Ratio_S: XX.X
KP: XX
rpm
KP_S: XX
rpm
END-Ratio: XX.X
END-Ratio_S: XX.X
B a l a n c e C h a m b e r
MSBK1: XXXXX
MSBK2: XXXXX
VEBK1_S: OPEN/CLOSED
VABK1_S: OPEN/CLOSED
VDEBK2_S: OPEN/CLOSED
VDABK2_S: OPEN/CLOSED
VEBK2_S: OPEN/CLOSED
VABK2_S: OPEN/CLOSED
VDEBK1_S: OPEN/CLOSED
VDABK1_S: OPEN/CLOSED
Flow (Dialysate Fluid)
FPE: XXXX
rpm
FPA: XXXX
rpm
FMD: XXX
ml/min
FMD_S XXX
ml/min
V a l v e s ( B y p a s s )
VBP: OPEN/CLOSED
VBP_S: OPEN/CLOSED
VDE: OPEN/CLOSED
VDE_S: OPEN/CLOSED
VDA: OPEN/CLOSED
VDA_S: OPEN/CLOSED
V a l v e s ( Disinfection)
VZ: OPEN/CLOSED
VD OPEN/CLOSED
VD_S OPEN/CLOSED
B l o o d L e a k
BL: X.XX
‰
BL_S: X.XX
‰
U l t r a f i l t r a t i o n
UFP: XX
rpm
UFP_S: XX
rpm
PDA: XXX
mmHg
PDA_S: XXX
mmHg
TMP: XXX
mmHg
B I C C a r t r i d g e H o l d e r
BKUS: OPEN/CLOSED
BKUS_S: OPEN/CLOSED
VVB: OPEN/CLOSED
VBICP: OPEN/CLOSED
VBKO: OPEN/CLOSED
VBKS: OPEN/CLOSED
V a l v e s O n l i n e
VBE: OPEN/CLOSED
VBE_S: OPEN/CLOSED
VDFF: OPEN/CLOSED
VDFF_S: OPEN/CLOSED
VSB: OPEN/CLOSED
VSB_S: OPEN/CLOSED
VSAA: OPEN/CLOSED
VSAA_S: OPEN/CLOSED
VSAE: OPEN/CLOSED
VSAE_S: OPEN/CLOSED
P u m p O n l i n e
OPS: XXXX
rpm
OPS_S: XXXX
rpm
Subst. Flow (Online)
FMS: XXX
ml/min
FMS_S: XXX
ml/min
Blood Side (Pressure)
PA: XXX
mmHg
PA_S: XXX
mmHg
PBE: XXX
mmHg
PV: XXX
mmHg
PV_S: XXX
mmHg
PBS: XXX
mmHg
PBS_S: XXX
mmHg
B l o o d P u m p s
BPA: XXXX
rpm
BPA_S: XXXX
rpm
BPV: XXXX
rpm
BPV_S: XXXX
rpm
T u b i n g C l a m p s
SAKA: OPEN/CLOSED
SAKV: OPEN/CLOSED
SAKV_S: OPEN/CLOSED
A i r D e t e c t o r
SAD: NO
AIR/AIR
SAD_S: NO
AIR/AIR
SAD TIME_S:
OK
SAD REF_S:
XXXX mV
SAD VOL:
XXXX µl
SAD VOL_S:
XXXX µl
H e p a r i n
HP: STOP
POSITION: XX.XX
mm
S t a t u s S e l f T e s t
LLC: XXXXX
LLS: XXXXX
L L C / S & T L C A l a r m W i n d o w
L L C M e s s a g e W i n d o w
Fig.: First Page Service Overview in Therapy
Balance Chamber
MSBK1: 32134
MSBK2: 32270
Digit Range for MSBK1/2 in Therapy
MSBK1/2
Min. Value:
10000
Max. Value:
35000
Differential Range:
min. 7500
Malfunctions in the balance chamber can occur if:
•
the values are out of limits
•
the minimal difference from the detected min. value was not reached