4.8.2.25
Test 1.25 HDF Online Valves and Coupling Detectors
1.
Activate the HDF Online Valves and Coupling
Detectors menu with the HDF Online Valves
and Coupling Detectors key in the LL Manual
Test menu. The following menu is displayed.
2.
Check the functions with the appropriate keys.
You can activate and deactivate the HDF online valves.
Valves
VBE, VDFF, VSB, VSAE, VSAA
You can either activate or deactivate a single valve with the Desired Setting
icon or all valves
VBE
,
VDFF
,
VSB
,
VSAE
,
VSAA
with the Common Setting icon.
Thus the feedback from the LLS can be checked.
Detectors
PSAUSS (substitution port outlet sensor)
The status of the additional coupling sensor
PSAUS
can be checked.
PSABFS (substitution port drain sensor)
The status of the additional coupling sensor
PSABFS
can be checked.
FEDFFS (filter recognition DF filter)
The status of the additional coupling sensor
FEDFFS
can be checked.
FEHDFS (filter recognition HDF filter)
The status of the additional coupling sensor
FEHDFS
can be checked.
PSAUS
PSABFS
FEHDFS
FEDFFS