B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.18
Stroke Value HDF Online Substitution Pump OSP
1.
Activate the LL Manual Calibration menu with
the LL Manual Calibration key in the Manual
Test and Calibration menu.
2.
Activate the HDF Online Substitution Pump
OSP menu by pressing on the line Calibration
2.15 HDF Online Substitution Pump OSP. The
following menu is displayed.
Stroke Value
HDF Online Substitution Pump OSP
The stroke value must be set to 8300 (for pump head 8 x 12).