4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 162
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.21
Calibration Dialysate Nonreturn
Valve RVDA
You can calibrate the dialysate nonreturn valve
RVDA
.
DDE
FBK2
FBK1
RVDA
VABK2
VDABK2
VABK1
VDABK1
BK2
BK1
VEBK2
VDEBK2
VEBK1
VDEBK1
MSBK2
MSBK1
VZ
RVFPE
FPE
8
Loop-In Point
for Calibration
Fig.: Calibration RVDA/Loop-In Point
Fig.: UF Sub-Rack with FBK1
Fig.: Water Sub-Rack with RVDA
FBK1
EP
FEP
RVDA
Loop-In Point
for Calibration
WAB