4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 15
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.5
Service Overview in TSM
Service Program
The service overview can be selected with the
key in the TSM service
program.
R e q u i r e d D a t a t o L L C
A c t u a l D a t a f r o m L L C
P U M P
BICP: X
BPA: X
BPV: X
EP: X
FPA: X
FPE: X
KP: X
UFP: X
H E A T E R
H: X
S A D
SAD_LEVEL: X
A C T I O N S
LEAKTEST: ON/OFF
EMPTYING: ON/OFF
BL_LAMP_T: ON/OFF
SAD_TEST: ON/OFF
V A L V E
RVE: X
VVBE: OPEN/CLOSED
VABK1: OPEN/CLOSED
VDABK1: OPEN/CLOSED
VEBK1: OPEN/CLOSED
VDEBK1: OPEN/CLOSED
VABK2: OPEN/CLOSED
VDABK2: OPEN/CLOSED
VEBK2: OPEN/CLOSED
VDEBK2: OPEN/CLOSED
VBP: OPEN/CLOSED
VDE: OPEN/CLOSED
VDA: OPEN/CLOSED
VD OPEN/CLOSED
VZ: OPEN/CLOSED
VLA: OPEN/CLOSED
VDFF: OPEN/CLOSED
VSB: OPEN/CLOSED
VBE OPEN/CLOSED
VSAE: OPEN/CLOSED
VSAA: OPEN/CLOSED
P U M P
BPA: X
BPAIMP_S: X
BPA_FLOW: X
BPA_COVER:OPEN/CLOSED
BPV: X
BPVIMP_S: X
BPV_COVER:OPEN/CLOSED
EP: X
FPE: X
FPA: X
UFP_S: X
UFP_FLOW: X.X
UFP_FLOW_S: X.X
BICP_S: X
BICP_FLOW: X.X
BICP_FLOW_S: X.X
KP_S: X
KP_FLOW: X.X
KP_FLOW_S: X.X
HP_POS: X.X
HPKOLB: ON/OFF
HPKRALLO: OPEN/CLOSED
P R E S S U R E
PA: X
PA_S: X
PV: X
PV_S: X
PBE: X
PBS: X
PBS_S: X
PDA: X
PDA_S: X
PE: X
L E V E L
NSVB: HIGH/LOW
LAFS: HIGH/MIDDLE/LOW
T E M P E R A T U R E
TSE XX.X
TSHE XX.X
TSBIC XX.X
TSD XX.X
TSD_S XX.X
TSDE XX.X
V A L V E
VBP_S: OPEN/CLOSED
BALANCE CHAMBER
MSBK1: XXXXX
MSBK2: XXXXX
VABK1_S: OPEN/CLOSED
VDABK1_S: OPEN/CLOSED
VEBK1_S: OPEN/CLOSED
VDEBK1_S: OPEN/CLOSED
VABK2_S: OPEN/CLOSED
VDABK2_S: OPEN/CLOSED
VEBK2_S: OPEN/CLOSED
VDEBK2_S: OPEN/CLOSED
C O U P L I N G S
BIC_SS: OPEN/CLOSED
K_SS: OPEN/CLOSED
SBS1: OPEN/CLOSED
SBS2: OPEN/CLOSED
BKUS OPEN/CLOSED
BKUS_S OPEN/CLOSED
C O N D U C T I V I T Y
BICLF: X.XX
ENDLF X.XX
ENDLF_S X.XX
AUTOMATIC TEST & CALIBRATION
LF: X.XX
PA: XXX
TSD: X.X
PDA:
XXX
TSE: X.X
W O R K I N G T I M E C O U N T E R
WTC: XXXXX
F R E E M E M O R Y ( M b )
XXXX of XXXXXX kB S: XX/XX MB
B L O O D L E A K
BLCON: X.XXX
BLCON_S: X.XXX
S A D
SAD_LEVEL: XX
SAD_LEV_S: X.XXX
SAD: AIR/BLOOD
SAD_S: AIR/BLOOD
VDE_S: OPEN/CLOSED
VDA_S: OPEN/CLOSED
VD_S OPEN/CLOSED
VZ_S: OPEN/CLOSED
VLA_S: OPEN/CLOSED
VDFF_S: OPEN/CLOSED
VSB_S: OPEN/CLOSED
VBE_S: OPEN/CLOSED
VSAE_S: OPEN/CLOSED
VSAA_S: OPEN/CLOSED
S T A T U S
RDV: NO
BLOOD/BLOOD
RDV_S NO BLOOD/BLOOD
P O W E R
B24VOFF: ON/OFF
D24VOFF: ON/OFF
L L C S T A T U S : 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
L L S S T A T U S : 0 0 0 0 0 0 0 0
Fig.: Service Overview in TSM Service Program
Balance Chamber
MSBK1: 32123
MSBK2: 32253
Digit Range for MSBK1/2 in TSM
MSBK1/2
Min. Value:
11000
Max. Value:
34000
Differential Range:
min. 5500
Malfunctions in the balance chamber can occur if:
•
the values are out of limits
•
the minimal difference from the detected min. value was not reached