4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 29
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.7.1.4
Arterial Bolus Parameters
1.
Activate the System Configuration menu with
the System Configuration key in menu
Treatment Support. Use the scroll bar until the
following menu appears.
2.
Activate the Arterial Bolus Parameters menu
with the Arterial Bolus icon. The following
menu is opened.
The following parameters can be set in the
Arterial Bolus Parameters menu:
•
Arterial Bolus Volume
•
Arterial Bolus with SAKA Support
Arterial Bolus Volume
The volume for an arterial bolus can be set.
Arterial Bolus with SAKA Support
The arterial tubing clamp SAKA can be used to support an arterial bolus in the
Arterial Bolus Parameters menu. Prerequisites are:
•
Arterial tubing clamp SAKA present
•
An arterial line system with infusion access is present between the SAKA
and blood pump