4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 21
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.7.1 System
Configuration
1.
Activate the System Configuration menu with
the System Configuration key in menu
Treatment Support.
The following menu is opened.
Scroll the various system configuration points up
or down with the scroll bar. System configuration
points can be selected (additional settings are
activated if specific options are selected).
4.7.1.1
Dialysate Side Parameters
The dialysate fluid parameters can be set in the
Dialysate Side Parameters menu.
1.
Activate the Dialysate Side Parameters menu
with the Dialysate Side Parameters icon. The
following menu is opened.
The following parameters can be set:
•
Conductivity measurement unit
•
Acetate Mode Disabled
•
Default Conductivity Mode
•
Stand-by Function is Available in Preparation
•
Stand-by after Self Test
•
Maximum Stand-by Time
•
Selected Acetate (only necessary if mmol/l is
selected)
•
Selected Bicarbonate/Selected Acid (only
necessary if mmol/l is selected)
•
Temperature Measurement unit
•
Dialysate Flow
•
Dialysate Temperature
Conductivity Measurement Unit
Make sure that the corresponding conversion factor is set and assigned
correctly:
•
conversion factor for the conductivity unit mS/cm
•
conversion factor for the concentration unit mmol/l.
You can either select conductivity [mS/cm] or sodium concentration [mmol/l]
for the conductivity measurement unit. If sodium concentration [mmol/l] is
selected a corresponding conversion factor must be entered. Activate the ACET
or BIC icon to enter the corresponding conversion factor.
1.
Click the mmol/l or mS/cm key to select a measurement unit.
Acetate Mode Disabled
The acetate mode can be disabled.
1.
Click the YES key to disable the acetate mode.