BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.2.17
Test 1.17 UF Pump
1.
Activate the UF Pump menu with the UF Pump
key in the LL Manual Test menu. The following
menu is displayed.
2.
Check the functions with the appropriate keys.
You can set the speed of the
UF
pump
UFP
. The head speed and the calculated
flow are displayed. At the beginning of the test the balance chambers are
initialised. The flow pumps
FPE/FPA
are activated to reach the inlet pressure
for
UFP
, set by the value of the throttle
DDE
.
Flow Detector
Not applicable.
Stroke Value UFP
The Stroke Value UFP should not be used in a normal case, because
UFP
must
be calibrated (see LL Manual calibration).