4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 78
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.2.4
Test 1.4 Heparin Pump
1.
Activate the Heparin Pump menu with the
Heparin Pump key in the LL Manual Test menu.
The following menu is displayed for the
heparin pump Compact.
2.
Check the functions with the appropriate keys.
HP Position
The forward feed distance is displayed in mm (only if the syringe is inserted
and locked). The counter is reset if the lock sensor is opened. The distance is
detected from the signal of the light barrier on the motor axle.
HP Syringe Sensor
The sensor for the sensor plate of the syringe detects an inserted syringe.
Lock Sensor
Sensor to open the clasp nut.
(The clasp nut is assembled on the drive spindle and limits the max. pressure
load of the heparin pump).
HP Motor State
Activates/switches off the heparin pump.
HP Speed
Selection between slow and fast forward feed.