4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 153
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
6.
Connect and check all concentrates in succession and note the
BICLF
ratio
and
ENDLF ratio
values.
7.
Change the treatment mode to Acetate and connect only acetate
concentrates if Acetate mode is performed in that dialysis centre.
8.
Wait for the conductivity to be stabilized and read the actual ENDLF ratio
from the service overview.
9.
Switch off the Dialog and set service switch S1 to position 2 TSM service
program.
10.
Switch on Dialog.
11.
Select the Manual Test & Calibration key.
12.
Select LL Manual Calibration key and activate 2.10 BICLF Ratio & ENDLF
Ratio.
For bicarbonate therapy with
BICLF ratio
values near the limits or out of the
default setting of 25 ±7.
11.
If necessary enter the value for
BICLF ratio
(mean value of all
bicarbonates) in BICLF Ratio Desired Setting Controller 25 key (default
value 25).
12.
Confirm and save the entered value with the Calibration key.
To prevent alarms during therapy in bicarbonate mode the value of the
ENDLF
Ratio
must be less than the set
ENDLF Ratio
e.g. <31.
To prevent alarms during therapy in acetate mode the value of the
ENDLF
Ratio
must be more than the set
ENDLF Ratio
, e.g. >31.
13.
If necessary select ENDLF Ratio - Desired Setting Controller - 31 (limit
Value 31) and enter the value for the
ENDLF ratio
(between highest value
for all acidic concentrates and lowest value for all acetate concentrates).
14.
Save the entered limit value.
15.
Switch off Dialog and set service switch S1 to position 0 therapy.
2. Bicarbonate Cartridge Holder Option
The option must be activated, set and checked in the TSM service program.
After the activation a ratio value must be set for the BICLF Cart Ratio. Check
the ratio value in therapy (service overview) and adjust if necessary.
S e t t i n g R a n g e : 25 to 50
L i m i t s :
Setting ± 10
1.
Enter 35 in Desired Setting Controller for the BICLF Cart Ratio.
2.
Save the entered value with the Calibration key.
3.
Check the bicarbonate ratio value in therapy.