B. Braun Avitum AG
Dialog
+
SW 9.xx
4.7.1.6 Heparin
Parameters
You can set the parameters in the Heparin
Parameters menu and select a syringe type.
1.
Activate the System Configuration menu with
the System Configuration key in menu
Treatment Support. The following menu is
opened.
2.
Activate the Heparin Parameters menu with
the Heparin Parameters icon. The following
menu is opened.
The following parameters can be set:
•
Treatment without heparin
•
Heparin measurement unit
•
Conversion factor heparin [IE/ml]
•
Heparin rate
•
Heparin bolus volume
•
Therapy beginning bolus
•
Selected syringe
Treatment Without Heparin
YES
The treatment is not started with heparin if the YES key is selected, i.e. the
default is set to treatment start without heparin. The heparin pump alarms are
disabled.
NO
The treatment is started with heparin if the NO key is selected, i.e. the default
is set to treatment start with heparin. The heparin pump alarms are enabled.
Heparin Measurement Unit
The measurement unit [IE/H] or [ml/h] can be selected for heparin.
Conversion Factor Heparin [IE/ml]
You can set the conversion factor ([IE/ML] depending on the mixing ratio, if the
amount of heparin should be displayed in international units (IE). The
conversion factor describes the quantity of IE in 1 ml solution, e.g. 1000 IE
represent in 1 ml solution 1000 IE heparin.
Heparin Rate
The heparin rate can be set to [IE/h] or [ml/h], depending on the selected unit.
Heparin Bolus Volume
You can select the bolus volume for the heparin bolus.
Therapy Beginning Bolus
An automatic bolus can be administered at the beginning of a therapy. The
heparin bolus is performed by the venous red detector RDV.