3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 94
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.6.13
Troubleshooting USB Stick
3.6.13.1 Field Service Utility Menu
with Failed CRC Check
The FSU program stops automatically in case of a self test error. The
corrupted USB stick must be exchanged before you continue with the FSU
program.
The USB stick can be damaged permanently if the procedure to remove the
USB stick is not observed.
R e q u i r e d C R C = X X X X
* * * S e l f = t e s t E R R O R – U S B s t i c k i s n o t u s a b l e * * *
- - - S W I T C H O F F t h e D i a l o g + - - -
The following error message is displayed in case of a
self test error.
Exchange the corrupted USB stick before you continue
with the FSU program.