4.7.1.20 DBI
Parameters
DBI parameters can be set in the Dialog Bedside-
Link Interface DBI Parameters menu.
1.
Activate the System Configuration menu with
the System Configuration key in menu
Treatment Support. The following menu is
opened and displayed (scroll with scroll bar).
The DBI menu is present and can be opened if DBI
Nexadia was selected in the Production Report,
Low Level Options. Various DBI parameters can be
selected in the DBI Parameters menu.
Baud Rate (kBaud)
The default for the Baud rate is 38.4 kBaud.
Automatic User Logout Time (min)
An automatic user logout time can be set and
activated. After the automatic logout the user has
to log in again before he is able to confirm any
messages, medications or checklist entries.
Message Warning Time (min)
A message warning time can be set and activated.
Thereby it is possible to inform users of any
unconfirmed messages, medications or checklist
entries, before the end of a therapy.
Note:
The data set must be set according to the
support of the Nexadia software.
Data Set DBI 1.61
(applicable for Nexadia)
Data Set DBI 1.72
(not applicable)
Data Set DBI 2.00
(default)
Data Set DBI 2.20
(not applicable)
Data Set DBI 2.21
The data set DBI 2.21 is required for the extension
of the Nexadia functionality, e.g. support of the
option bioLogic RR Comfort and Adimea; the
mmol/l unit can be used. The Nexadia protocol
must be supported by the host computer.