4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 159
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.19 Level
Regulation
1.
Activate the LL Manual Calibration menu with
the LL Manual Calibration key in the Manual
Test and Calibration menu.
2.
Activate the Level Regulation menu by
pressing on the line Calibration 2.16 Level
regulation. The following menu is displayed.
Level Regulation
The speed of the level regulation pump (LRP = PPR) can be set for the slow
mode and for the fast mode.
The settings are (see default table in chapter 3):
•
60 % for the slow mode
•
100 % for the fast mode
The default settings should not be changed, unless instructed otherwise.