4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 143
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.10.2 Check Blood Leak Detector
1.
Connect both dialyser couplings to the
rinsing bridge.
2.
Select
menu
Test 1.18 Blood Leak in TSM to
perform a function check of the blood leak
detector.
3.
Rinse the blood leak detector for approx. 1
min at the beginning of the rinsing time
with flow pump outlet
FPA
(approx. 1400
rpm). Thereby possible air bubbles in the
system are removed.
4.
Rinse blood leak detector for approx. 15 to
30 min with
FPA
(approx. 1000 rpm) to
warm up the complete system to a
temperature of approx. 40
o
C.
Exchange the blood leak detector if one of these three values is out of
limits.
5.
The
LED test is activated with the LAMP TEST OFF key. The OFF key
switches to the RED key as soon as the key is pressed.
The
red LED is tested after 10 s. The controller value for the switched off
red LED must be as follows (the supervisor value is uncritical):
red LED
Actual Status Controller
Blood Concentration
< 4.00
6.
Press
the
LAMP TEST RED key to switch to the green LED. The RED key
switches to the GREEN key.
The green LED is tested. The controller value for the switched off green
LED must be as follows (the supervisor value is uncritical):
green LED
Actual Status Controller
Blood Concentration
> 2.06
7.
The actual controller values for the red LED and the green LED must be
divided.
Note:
The calculated value of
red LED
divided by
green LED
is displayed
briefly after the key LAMP TEST is switched back to OFF.
The calculated value must be in the following limits (to the first decimal
digit/round off):
red LED / green LED
1.15 - 1.44