3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 68
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
BlueCat
Removable Disk (X: - FSU USB Stick)
DISKTYPE
USER
05_Write_Calibration_Data
DEFAULT
04_Write_Languages
The utility searches for input in the 05_Write_Calibration_Data, DEFAULT
subdirectory
and
in the results of the previously performed 11. Read Calibration
Data utility. If no file was found the following message is displayed:
No source…
If there are files (e.g. read from other machines) but there are no
matching files, the following message is displayed:
No selectable source is found…
and the existing serial number/system version number pairs are displayed with
the following message:
(not selectable)
If there is one matching file the following message is displayed:
Are you sure?
If there are more matching files (for the calibration data this is only possible if
you read the calibration data with the 11. Read Calibration Data utility and
copy the result file manually to the 05_Write_Calibration_Data, DEFAULT
subdirectory): all existing serial number/system version number pairs are
displayed (DEFAULT is displayed instead of the serial number if the subdirectory
contains a calib.dat.xxu file). The user can select from the matching pairs only.
The pairs that do not match are displayed with the following message:
(not selectable)
5.
Press
EQ
if you are sure. This will write the calibration data from the
USB stick to the hard disk drive (CFC).