4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 18
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.6 File
Operations
The File Operations menu is selected with the File
icon in TSM Main Menu.
The following parameters can be selected:
Floppy Diskette Read/Write/Create
Not applicable.
Hard Disk (Compact Flash Card) Read/Write
•
Save Calibration Data
•
Down Load Calibration Data
The Field Service Utilities FSU can be used for all relevant service
activities. Details can be found in chapter 2 repair instructions.
4.6.1
Hard Disk Read/Write
Save Calibration Data
The calibration data must be saved to the compact flash card CFC (hard disk
drive) by LLC/LLS after every calibration.
Down Load Calibration Data
The calibration data can be downloaded from the compact flash card CFC (hard
disk drive) to the exchanged digital board DB.