4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 28
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Standard TMP Limit Range
-100
+300
Top TMP Limit Value
Bottom TMP Limit Value
Relative Limits
at 40 %:
30 - 70 mmHg
Current
TMP
40
30
60
70
50
min. TMP Alarm Window
20 mmHg
0
TMP
[mmHg]
-10
Example A
Top TMP Limit Value
Bottom TMP Limit Value
Relative Limits
at 40 %:
12 - 28 mmHg
12
28
20
min. TMP Alarm Window
20 mmHg
TMP
[mmHg]
10
30
-100
+300
0
-10
Current
TMP
Example B
If the Standard TMP Limit Range key is not
pressed:
•
the TMP low limit range is set min. –10
mmHg.
If the normal window limits are reached:
•
a warning message is displayed (in alarm
window)
•
a warning tone is activated.
•
treatment is not interrupted
Once the TMP reaches -10 mmHg:
•
a warning message is displayed (in alarm
window)
•
a warning tone is activated
•
an additional information is displayed
•
treatment is not interrupted
The absolute limit value is -100 mmHg.
Once the absolute TMP value is reached:
•
an alarm is activated
•
an alarm tone is activated
•
treatment is interrupted
•
machine is switched to bypass
In example
A
the min. TMP alarm window (40 to
60 mmHg) is smaller than the relative limits of 30
to 70 mmHg. Thus the relative limits are valid.
In example
B
the min. TMP alarm window (10 to
30 mmHg) is larger than the relative limits of 12
to 28 mmHg. Thus the limits of the min. TMP
alarm window are valid.
Extended TMP Limit Range
Top TMP Limit Value
Bottom TMP Limit Value
Relative Limits
at 40 %:
12 - 28 mmHg
12
28
20
min. TMP Alarm Window
20 mmHg
TMP
[mmHg]
10
30
-100
+300
0
Current
TMP
If the Extended TMP Limit Range key is pressed:
•
the TMP limit range is extended to –100
mmHg (absolute TMP limit).
Once this absolute TMP limit is reached:
•
an alarm is activated
•
an alarm tone is activated
•
a warning message is not displayed anymore
•
treatment is interrupted
•
machine is switched to bypass