Test 1.16 Pressure Water Part
1.
Activate the Pressure Water Part menu with
the Pressure Water Part key in the LL Manual
Test menu. The following menu is displayed.
2.
Check the functions with the appropriate keys.
You can skip directly to the calibration of the pressure sensors
PE
and
PDA
by
pressing the
key.
The degassing pump
EP
can be set. The degassing bypass valve
VEB
can be
opened/closed.
The pressure at the degassing pressure sensor
PE
can be set with the degassing
pump
EP
and the degassing bypass valve
VEB
.
The pressure at the outlet dialysate pressure sensor
PDA
is set manually with a
syringe at the dialyser couplings.
Pressure in mbar
The pressure can be displayed in mmHg or in mbar, depending on selection.
Conversion Factor mmHg to mbar
Conversion Factor = 1.3332
Example:
-100 mmHg
approx. –133 mbar
–400 mmHg
approx. –533 mbar