3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 97
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Unit
HD
HDF Online
US Default
Modified
HDF/HF Online Parameters
HDF Substitution Flow
ml/min
-
60
-
HF Substitution Flow
ml/min
-
100
-
HDF Dialysate Flow for Postdilution
ml/min
-
600
-
HDF Dialysate Flow for Predilution
ml/min
-
700
-
HDF/HF Infusion Bolus Volume
ml
-
100
-
Post-/Predilution Selection
-
-
inactive
-
Disinfection Parameters
Disinfection after each Therapy
-
NO
YES
NO
Automatic Preparation Start after Disinfection
-
NO
NO
NO
Thermal Disinfection
-
YES
NO
YES
Chemical Disinfection
-
YES
YES
YES
Chemical Disinfection, Short
-
YES
NO
NO
Central Thermal Disinfection
-
NO
NO
NO
Central Manual Chemical Disinfection
-
NO
NO
NO
Central Automatic Chemical Disinfection
-
NO
NO
NO
Rinsing
-
NO
NO
YES
Automatic Switch-On with Rinsing - No Weekly Schedule
-
inactive
inactive
inactive
Switch-On Time
h:min
06:30
06:30
06:30
Automatic Switch-On by Weekly Schedule
-
active
active
active
Weekly Schedule
(see Table)
Start on Sunday
-
inactive
inactive
active
Start on Monday
-
active
active
inactive
Auto Switch-On without daily confirmation
-
NO
NO
NO
Maximum Out of Action Time
-
inactive
inactive
active
Days
Days
0
0
2
Hours/Minutes
h:min
08:00
08:00
00:00
Disinfection Configuration Data Table
-
Citric Acid 50%
Citric Acid 50%
Citric Thermal
Central Thermal: Inlet Flow
ml/min
250
250
250
Central Thermal: Time
h:min
00:30
00:30
00:30
Central Manual Chemical: Inlet Flow
ml/min
250
250
250
Central Manual Chemical: Inlet Time
h:min
00:05
00:05
00:05
Central Manual Chemical: Rinsing Flow
ml/min
800
800
800
Central Manual Chemical: Rinsing Time
h:min
00:30
00:30
00:30
Central Automatic Chemical: Inlet Volume
ml
260
260
260
Central Automatic Chemical: Switch off for Retention
without Automatic Switch-On
-
NO
NO
NO
Central Automatic Chemical: Retention Time
h:min
00:20
00:20
00:20
Central Automatic Chemical: Rinsing Flow
ml/min
800
800
800
Central Automatic Chemical: Rinsing Time
h:min
00:30
00:30
00:30
Rinsing: Inlet Flow
ml/min
800
800
800
Rinsing: Time
h:min
00:02
00:02
00:02
Filter Parameters
Name
-
Diacap-Ultra
Diacap-Ultra
Diacap-Ultra
Therapy Number
-
150
150
150
Filter Operation Time
h
900
900
900
Dialyser Parameters (Kt/V)
Urea Distribution Volume Formula
-
Watson
Watson
Watson
Height Measurement Unit of Watson Formula
-
cm
cm
cm
Default Target Kt/V
-
1.20
1.20
1.20
Default Setting for Treatment Mode Warning of Kt/V
Target Projected Deviation
-
Warning active
Warning active
Warning inactive
Kt/V Table - Manual Input: Setting Clean/Total Blood [%]
-
YES
YES
YES
Warning for Saving Kt/V Volume enabled
-
NO
NO
NO
Kt/V-UV
Default Target Kt/V
-
1.20
1.20
1.20
Kt/V Correction Mode
-
Single Pool
Single Pool
Single Pool
Default Setting for Treatment Mode:
Warning of Kt/V Target-Projected Deviation
-
Warning active
Warning active
Warning inactive
Warning for Saving Kt/V Table
-
NO
NO
NO
Timings
Parameter Setting Window Disappearing Time
s
500
500
600
Screen Saver Appearance
-
NO
NO
NO
Appearing Time
s
600
600
600