4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 110
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.2.31 Test 1.31 Level Regulation
1.
Activate the Test 1.31 Level Regulation menu
with the Level Regulation key in the Low Level
Manual Test menu. The following menu is
displayed.
2.
Check the functions with the appropriate keys.
PA
PV
PBE
PBS
1
2
3
1
1
1
2
2
2
3
3
3
PA Chamber
PV Chamber
PBE Chamber
HFA
HFV
HFE
HFS
VBT
VPV
VPE
HFP
VPU
LRP
VPD
VPA
Front Door
Inside
Outside
Level Module
1
1
2
2
3
3
P
OUT
P
IN
The valves VPE, VPD, VPU, VPV, VPA, and VBT are
all assembled on the level regulation module. All
valves can be activated and tested. The
manometer connectors must be opened for the
test.
Actual Status Supervisor
LRP (PPR) OFF/ON
The supervisor must detect correctly the current
status of the level regulation pump LRP (PPR).
Desired Setting
OFF/ON
The level regulation pump LRP (PPR) can be
switched on and off.
Slow/Fast
The desired value for the LRP can be switched
between a Slow and a Fast setting.
Test
The delivery of the LRP can be checked with the
Test key. A measurement chamber must be
connected to PV. LRP stops when 500 ml are
reached and the required time for the pressure
build-up is displayed.
VPE VPD
VPU
VPV VPA
VBT
LRP
HFS
HFE
HFV
HFA
PA
PBS
PV
PBE
HFP
VPE
VPD
VPU
VPV
VPA
VBT