3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 98
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Unit
HD
HDF Online
US Default
Modified
Language Selection
English
-
active
active
active
German
-
inactive
inactive
inactive
Summer Time Setting
(enter values accordingly)
Begin of Summer Time
D/M/Y
dd.mm.yyyyy
dd.mm.yyyyy
dd.mm.yyyyy
h:min
xx:xx
xx:xx
xx:xx
End of Summer Time
D/M/Y
dd.mm.yyyyy
dd.mm.yyyyy
dd.mm.yyyyy
h:min
xx:xx
xx:xx
xx:xx
Time Difference
h:min
xx:xx
xx:xx
xx:xx
ABPM
Delete Button for Data
-
YES
YES
YES
Auto Start of Cyclic Reading (from Patient Identification
Media)
-
NO
NO
NO
ABPM Cyclic Reading Stop at Entering Disinfection
-
YES
YES
YES
Taking Over Limits from Patient’s Parameters
-
YES
YES
YES
bioLogic RR Comfort
Guide Line Mode
-
NO
NO
-
Suggested SYS Lower Limit
-
NO
NO
-
bioLogic RR Algorithm ID
-
100
100
-
Miscellaneous Parameters
Click Sound
-
NO
NO
YES
Warning and Writing (Saving) Patient Parameters
-
NO
NO
YES
* Skip Self Tests
* A skip of the self tests for servicing only!
-
NO
NO
NO
Automatic Preparation Start
-
YES
YES
NO
Data Validation with
-
Touch
Touch
Enter Key
UF Profile Editing in Treatment
-
YES
YES
NO
Manual Bypass Confirmation Window Enabled
-
NO
NO
YES
End of Therapy Sound Duration
-
100 %
100 %
100 %
Chopped Alarm Sound
-
NO
NO
NO
Suppression of Warning Sounds in Preparation
-
NO
NO
NO
Automatic Reinfusion Start at Entering End of Therapy
-
YES
YES
NO
DCI
Dianet Address
-
1
1
1
Data Set
-
MPI 0.12
MPI 0.12
MPI 0.12
DBI
Baud Rate
kBaud
38.4
38.4
38.4
User Logout Time
min
15 - ON
15 - ON
15 – ON
Message Warning Time
min
15 - OFF
15 - OFF
15 - OFF
Data Set
DBI 1.61
-
inactive
inactive
inactive
DBI 1.72
-
inactive
inactive
inactive
DBI 2.00
-
active
active
active
DBI 2.20
-
inactive
inactive
inactive
DBI 2.21
-
inactive
inactive
inactive
DBI-WAN Interface – Network Name Configuration
Pre-SN String is needed
-
-
-
d / YES
SN String is needed
-
-
-
- / YES
Post-SN String is needed
-
-
-
/ NO
Network Name (max. 15 chars.)
-
-
-
d-
Preventive Maintenance
Warning for Preventive Maintenance Enabled
-
NO
NO
NO
Load Interval
-
inactive
inactive
inactive
Elapsed Month
-
inactive
inactive
inactive
Setting Elapsed Month
months
12
12
12
Elapsed Working Time Counter WTC
-
inactive
inactive
inactive
Setting Elapsed WTC
hours
4000
4000
4000