4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 23
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Selected Bicarbonate/Selected Acid
There are 20 concentrate names and the
conversion factors for the conversion of
conductivity values.
If necessary the name and the conversion factor
[mS/cm per mmol/l] can be set, for the selected
bicarbonate type.
1.
Click the BIC icon to activate the menu for:
- BIC without NaCl
- BIC with NaCl.
BIC without NaCl
The current bicarbonate type without NaCl can be
selected:
•
Bicarbonate without NaCl (2 - 4 mS/cm)
Activate BIC without NaCl by pressing a free
section of the window.
1.
Click the BIC without NaCl window if
necessary to edit the bicarbonate name.
2.
Click the Conversion window if necessary to
change the conversion factor [mS/cm per
mmol/l].
3.
Click the Default Value window if necessary to
change the value [mS/cm].
4.
Click the ACID icon if necessary to edit the
following parameters:
•
Acid Concentrate Name
•
Conversion Factor [mS/cm per mmol/l]
•
Default Value