4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 37
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Note
If both the LIN and EXP keys are not activated the response of the graph can be
selected individually. If an interval is changed the remaining intervals are
adapted accordingly. The last interval can not be modified, because this setting
is calculated from the previous interval settings.
Current Profile No.
The profile selected with this key is displayed and can be modified. Additionally
to 10 standard profiles 20 further profiles can be activated, displayed and
modified.
Profile No. 0
This is a constant 100% profile.
Profile No. 1 to 9
These profiles are pre-defined.
Profile No. 10
A profile which was edited in therapy and stored to a chip card (card reader
option).
Profile No. 11 to 30
These profiles can be edited in the TSM service program and stored under a
number.
UF Mode HD
The UF process of the activated interval is set to HD with the HD key.
UF Mode Sequential SEQ
The UF process of the activated interval is set to sequential with the SEQ key.