Chapter 3—Setting Treatment Parameters
86
2008K Hemodialysis Machine Operator’s Manual Rev P
Table 19 – Control Panel Heparin on/off Key
Key
Function
The
Heparin on/off
key activates the heparin administration program.
It is located on the lower, right side of the control panel. The green
light above the button is illuminated when the pump is running.
When the light is flashing, the operation is interrupted and will resume
when the Treatment Clock is turned on or the
Heparin on/off
key is
pressed.
If the
Heparin on/off
key is pressed to turn off the pump, the light will
turn off and the pump will not resume with the
Treatment Clock
button.
The Heparin Delivery System
Warning!
The correct syringe type must be selected to ensure an accurate infusion.
To prepare the heparin delivery system using the features on the “Heparin” screen:
Figure 41 – The Syringe Subscreen on the Heparin Delivery Screen
1) Fill the
syringe selected
from the menu
with the
prescribed
amount of
heparin for the
entire treatment.
4) Press
CONFIRM
.
3) Use the Up/Down
keys on the data
entry keypad to
scroll the drop-down
menu until the
correct syringe type
is highlighted.
2) In the “Heparin”
screen, touch the
Syringe
button.
The syringe menu
opens.