2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
Set Venous Limits
This is a message to adjust the
venous limits if necessary
Advisory
Press and hold
RESET
for 1 second to adjust the limits. Changes in
venous pressure during the treatment should be investigated. See
“Venous Pressure Alarm”
Short Power Down
The machine was switched off for 1 –
2 minutes and turned back on. Setup
values have not been set to default
values.
Advisory
Verify that the dialysis parameters are as desired.
Single Needle On!
An action has been initiated that
requires the Single Needle option to
be off.
Warning
To proceed with the selected operation, de-select the Single Needle
option.
SN BP +5 V Error
+ 5 volts is outside the allowable
range
Alarm
See message E.10
SN BP +12 V Error
+ 12 volts is outside the allowable
range
Alarm
See message E.07
SN BP -12 V Error
- 12 volts is outside the allowable
range
Alarm
See message E.09
SN BP +24 V Error
+ 24 volts is outside the allowable
range
Alarm
See message E.08
SN BP Button Alarm
Key stuck or held in too long
Alarm
See message A.16
SN BP Comm. Timeout
Time out when receiving Intel-Hex-
line or overflowed received buffer
Alarm
See message A.27
SN BP Del. Rate Alarm
Actual speed-read back analog
voltage at X348/10 is out of limits
Alarm
See message A.21
SN BP Direction Alarm
Pump is turning in the wrong direction
Alarm
See message A.13
SN BP EEPROM Error
EEPROM error
Alarm
See message E.05
SN BP EPROM Error
EPROM CRC error
Alarm
See message E.01
SN BP Erasing Error
Error erasing Flash ROM while in
Service Mode
Alarm
See message E.98
SN BP Flash Error
Error copying data into Flash ROM
while in Service Mode
Alarm
See message E.97
SN BP Level Up Alarm
Pressure increase when the Level Up
key is pressed
Alarm
See message A.25