Chapter 4—Monitoring the Treatment
2008K Hemodialysis Machine Operator’s Manual Rev P
111
Feature
Function
The value displayed here is the same value entered in the
Target Kt/V
button in the “Kt/V AF” screen.
This data box displays the current calculated amount of single
pool Kt/V delivered therapy.
This data box displays the calculated equilibrated Kt/V. It is
calculated one hour after the beginning of treatment. The box
remains blank until then.
This data box displays the value for the equation (time
weighted mean Kecn) x (current time).
This is the result of the Access Flow test. It is limited to <2000
ml/min.
Data Table
—This table displays the individual and mean
Kecn data for the OLC and Access Flow tests.
Time refers to when the tests was performed in respect to
amount of time (hours:min) elapsed from the beginning of
treatment.
Manual tests are preceded with “M”.
Tests done for Access flow with the lines reversed are
preceded with “AF”. These tests are not used in the Mean
Kecn value.