154
2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
BPM: Cuff Press High
Blood pressure cuff is above 320 mm
Hg for adult or above 220 mm Hg for
pediatric mode.
Blood
Pressure
Alarm
Press the
Stat/Deflate
key to deflate the cuff. Observe the patient for
physiologic changes. Indicates a hardware malfunction. If problem
persists, remove cuff and alert a qualified service technician.
BPM: Cuff Press Low
Blood pressure cuff is below 10 mm
Hg.
Blood
Pressure
Alarm
Check for loose connection in the inflation system. Indicates a
hardware malfunction. If problem persists, remove cuff and alert a
qualified service technician.
BPM: Diastolic High
The diastolic blood pressure reading
is above the set Upper Diastolic alarm
limit.
Blood
Pressure
Alarm
Observe the patient for physiologic changes. Treat patient as
symptoms warrant.
BPM: Diastolic Low
The diastolic blood pressure reading
is below the set Lower Diastolic alarm
limit.
Blood
Pressure
Alarm
Observe the patient for physiologic changes. Treat patient as
symptoms warrant.
BPM: Measure > 90 sec
The blood pressure test has been in
progress for more than 90 seconds.
Blood
Pressure
Alarm
Press
Stat/Deflate
key to deflate the pressure cuff. Check the
patient for signs for physiologic changes. Indicates a hardware
malfunction. If problem persists, remove cuff and alert a qualified
service technician.
BPM Motion Detected
Movement of the patient, cuff tubing,
or some other pressure on the
detection system.
Blood
Pressure
Alarm
Observe the patient for physiologic changes. Treat patient as
symptoms warrant.
BPM: Not Communicating Blood pressure module is not
communicating with the machine
Blood
Pressure
Alarm
If problem persists, alert a qualified service technician.
BPM Not Deflating
Obstruction in inflation system or
valve in blood pressure module
malfunction.
Blood
Pressure
Alarm
1) Remove kink in line to cuff.
2) May indicate a hardware malfunction. If problem persists,
remove cuff and alert a qualified service technician.
BPM: Oscil Wave Check
The diastolic blood pressure reading
is close to or greater than the systolic
pressure reading.
Blood
Pressure
Alarm
Indicates a hardware malfunction. If problem persists, remove cuff
and alert a qualified service technician.
BPM: Pulse > 100
Patient’s heart rate is above 100
beats per minute.
Blood
Pressure
Alarm
Observe the patient for physiologic changes. Treat patient as
symptoms warrant. May also indicates a hardware malfunction.